Save up to 80% on pre-owned Baxter Travenol 6301 infusion pumps.
Request a Quote for this product now.
Product Specifications
32.5 cm W x 13 cm D X 29 cm H (13 in W x 5.1 in D x 11.4 in H)
12.0″ x 14.75″ x 7.5″ (30.5 cm x 37.5 cm x 19.1 cm)
8.1 Kg (17.9 lb)
12 volt, 2.0 ampere, sealed lead acid
80% charged within 8 hr.
6 hr at 1-1,400 mL/hr (4 hr with both channels operating at 1-1,400 mL/hr)
Biomed replaceable 2.75 m (9 ft)
Less than 50 micro-amps (per UL 544)
.8 ampere, 250 volt, slow-blow
RS232 serial port
15.5-37.7ºC (60-100ºF)
115 VAC, 60 HZ, 50 W
Primary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-1999 mL/hr in 1 mL increments.Upper limit can be reduced by authorized service personnel.
Secondary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-999 mL/hr in 1 mL increments
1.0-99.9 mL in 0.1 mL increments or 1-9999 mL in 1 mL increments for both primary and secondary. Upper limit can be reduced by authorized service personnel.
Total VTBI: 100-9,999 in 1 mL increments. Upper limit can be reduced by authorized service personnel.
Delay Time: 00:00 – 08:00 hours in minute increments.
Run Time: 01:00 – 16:00 hours in minute increments.
Ramp Up: 00:00 to 45% of the Run Time, in minute increments.
Ramp Down: 00-00 to the Run Time minus Ramp Up, in minute Increments.
Air-in-Line Detection:
Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger. The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.