Alaris/Cardinal IVAC Signature Gold 7130/7230 Infusion Pump




Get A Quote

Save up to 80% on pre-owned Alaris/Cardinal IVAC Signature old 7130/7230
infusion pumps. Request a Quote for this product now.

Product Specifications

Flow Rate Range:

0.1 to 999.9 ml/hr in 0.1 ml/hr increments (primary)
0.1 to 270.0 ml/hr in 0.1 ml/hr increments (secondary)

Volume To Be Infused Range:

.1 to 9999.9 ml in 0.1 ml increments (primary)
.1 to 999.9 ml in 0.1 ml increments (secondary)

Volume Infused Range:

0.0 to 9999.9 ml in 0.1 ml increments

KVO Flow Rate:

0.1-20.0 ml/hr in 0.1 ml/hr increments

Alarms:

Air in Line
Battery Depleted
Channel Malfunction
Computer Link Failure
Flow Sensor Unplugged
Hold Time Exceeded
Instrument Malfunction
Key stuck
Latch Open
No Upstream Flow Detected
Occlusion Downstream
Occlusion Upstream
Primary Flow Detected during

Secondary:

Set Out
Set Up Time Exceeded

Dimensions

7130:

Width 7.7 in/19.3 cm
Height 8.6 in/21.8 cm
Depth* 5.0 in/12.7 cm
Weight 6.85 lbs./3.11 kg
Power Cord 10 ft/3 m

7230:

Width 10.5 in/26.7 cm
Height 8.6 in/21.8 cm
Depth* 5.0 in/12.7 cm
Weight 8.4 lbs./3.7 kg
Power Cord 10 ft/3m

Case:

Impact resistant plastic

Power Requirements:

100- 120 V~,50/60 Hz, 0.5A,3-wire grounded system (40 watts Max)

Ground Current Leakage:

Tested to UL Standard 544 and CSA C22.2 No. 125 for medical and dental equipment.

Battery:

Rechargeable nickel cadmium. With a new, fully charged battery, the pump will operate for 4 hours nominal at 100 ml/hr for a two channel instrument operating on both channels simultaneously.

Environmental Conditions

Operating:

Temperature Range 10ºC to 40ºC
(50ºF to 104ºF)
Relative Humidity 15 to 95% Non-condensing
Atmospheric Pressure 632 to 1031 mbar

Storage:

Temperature Range -40ºC to 60ºC (-40ºF to 122ºF)
Relative Humidity 5 to 95%
Atmospheric Pressure 632 to 1031 mbar

For service, parts, PM or rental see our Services section